SONIMIX 4001
Ozone generator – Primary transfer standard
The Sonimix 4001 is a stand alone ozone primary transfer standard able to generate ozone together with dry or wet air (optional) allowing to calibrate the ozone monitors in several points. It generates an outflow of 3 Nl/min air containing 10 to 1000 ppb O3, with stability and reproducibility of the concentration better than 1% relative. The extended range model generates 10 ppb to 10 ppm at 3 L/min and 10 ppm to 35 ppm at 0.9 L/min.
The Sonimix 4001 is using pressure regulator and sonic nozzles and an ozone generator including thermostatisation and regulation by UV light photometer measurement with atmospheric pressure and temperature corrections (US EPA method of UV absorption by ozone, Beer-Lambert law). The ozone analyser is built with 1 reference UV light detectors that measure in alternance the UV intensity without any ozone and then the same UV light but through the generated ozone. With these 2 values, the processor calculates the amount of ozone generated.
Product features
- Stable and accurate > 1% (rel.)
- Calibration with dry & wet air
- Included novel LED photodetector
- Ergonomic with LCD display
Main applications
- Air pollution monitoring
- Laboratory (ozone transfer)
Technical specifications
-
Models:
Sonimix 4001
6800 17 000
standard range
6800 17 001
extended range
General data Range
From 10 to 1000 ppb in 1 ppb steps @ 3 L/min
From 10ppb to 10 ppm @ 3 L/min + 10ppm to 35 ppm @ 0.9 L/min
Dilution principle
Sonic nozzles and high precision pressure regulators
Dilution air
Dry air from internal zero air (option) or external dry air source
Generated gas flow
3000 ml/min for range 1
900ml/min for range 2
Accuracy
10 ppb to 100 ppb : +/- 2 ppb
>100 ppb : 1.5% relative
Ozone Repeatability
Better than 1% relative
Warm up time
< 45 min
Stabilization time
< 15 min
O3 regulation
Measure of the UV light absorption of the generated ozone by UV reference photometer (US EPA method of UV absorption by ozone, Beer-Lambert law). Including atmospheric pressure and room temperature corrections.
Dimensions
19cmx36cm, 50cm deep
Net weight
15 Kg
UV Section
UV lamp life time
About 10’000 hours (with a mean concentration of 1 ppm)
Zero Air section
Life Time
Annual preventive maintenance
Quality
O3, NO and NO2 < 1 ppb – H2O < -40°C dew point
Communication
Outlet Port
Teflon 1/4‘’ Swagelok fitting
Electrical data
Power supply voltage
230V/50-60Hz – 117V/60Hz – 100V/60Hz
Which LNI Swissgas solution is best suited for calibrating gas analyzers, gas sensors, or gas detectors?
LNI Swissgas offers several calibration solutions depending on the application.
The Sonimix 2106 and Sonimix 7100 generate accurate gas concentrations through dynamic dilution from gas cylinder. While the Permacal generates trace-level concentrations of reactive compounds such as NH₃, HCl, HF, HBr, VOCs, and hydrocarbons from permeation tubes.
Together, these solutions cover a wide range of gas analyzer, sensor, and gas detector calibration applications.
How to calibrate gas analyzers within an air monitoring station?
The Sonimix 6000 (C1 and C2) and Sonimix 3022/3012 are all-in-one multigas calibrators that integrate a gas mixing module, an ozone generator and a zero-air generator in a single compact enclosure.
They provide a complete solution for the calibration of SO₂, O₃, NOx, CO, and hydrocarbon analyzers. By generating zero air internally, these instruments eliminate the need for an external carrier gas supply, simplifying installation and reducing operating costs.
Calibrator or gas mixer: which one should you choose based on your application?
The term “calibrator” is often used when the primary application is the calibration of a gas analyzer or sensor. The term “gas mixer” is more commonly used when the objective is to generate a specific gas mixture.
Generating a gas mixture can serve different purposes, such as supplying an instrument, simulating a process gas, or calibrating an analyzer or sensor. In many cases, a gas mixer and a calibrator are essentially the same device, with the terminology depending mainly on the intended application.
Is it suitable for audit or official regulatory inspection environments?
Yes, all systems can be linked to the national standards.
Does the system require periodic maintenance or frequent recalibration?
The Sonic Nozzle technology provides unmatched stability, resulting in minimal drift and a recommended service interval of two years.
Does the system reduce logistics and gas storage costs?
Yes, our gas mixer helps reduce the number of gas cylinders required. Using a single high-concentration gas cylinder, the gas mixer can generate multiple concentration levels by dilution with a carrier gas.
This reduces the number of cylinders needed, lowers costs, and simplifies the associated logistics and cylinder management.
Can the system operate in continuous mode (24/7)?
Some customer are running our system in their production (24/7), no problem.
Is the ozone generator pressure – & temperature – compensated?
Yes, correct the Sonimix 4001 with atmospheric pressure and temperature corrections (US EPA method of UV absorption by ozone, Beer-Lambert law).
Can solenoid cylinder switching valves be directly controlled via the software?
Yes, all valves within the system are controlled automatically by the software.
Does the software log all events (flows, cylinder changes, errors) to ensure traceability and audit compliance?
Yes, the log allows to track the operational events.
Is it possible to program automatic multi-point calibration sequences?
Yes, all devices include a sequence mode which allows to build a calibration sequence to automatize a calibration process.
Is an accredited calibration certificate provided for the entire system?
LNI Swissgas is accredited according to ISO/IEC 17025 for flow measurement calibrations. An ISO/IEC 17025 calibration certificate can be supplied with every device.